System control: Equipment
2-13
C O N F I D E N T I A L
– limited rights
User Manual
Revision A
Feb 2018
Equipment
Detector types and application
There are several either standard or optional system detectors. The
Detectors
menu
shows all installed detectors. The accessibility / inaccessibility of detectors (black /
gray label) depends on actual system conditions.
Note
For settings and handling of particular standard / optional detectors, see Chapters 5 and 7.
SE = secondary electrons, BSE = back scattered electrons, TE = transmitted electrons,
S = standard, O = optional, LM = lens mounted version
Note
In case the vacuum safety interlock is activated, it is necessary to confirm switching the detector voltage on after pumping
the system.
Table 2-2:
Apreo detectors list
Detector name
Tag
Vacuum mode
Detected signal
Note
Everhart-Thornley
ETD
HiVac
SE (tunable energy) / BSE
S
T1
T1
HiVac / LoVac
SE (for OptiPlan and Immersion
Use cases)
S
T2
T2
HiVac / LoVac
SE (for OptiPlan and Immersion
Use cases)
S
Low Vacuum
LVD
LoVac
SE
S
CCD camera
CCD
light, infra-red light
S
External
EXT
detector-dependent
T3
T3
HiVac
SE (for OptiPlan and Immersion
Use cases)
O
RGB Cathodeluminescence
CLD
HiVac / LoVac
Light
O
Directional Gaseous
Analytical
GAD–ABS /
GAD–CBS
LoVac
BSE
O, LM
Directional Backscattered
(DBS)
ABS / CBS
HiVac / LoVac
BSE
O, LM
Retractable Directional
Backscattered (DBS)
ABS / CBS
HiVac
BSE
O
Retractable Annular STEM
STEM 3
STEM 3+
HiVac
TE
O