2
Installation
Placement and Environmental Requirements
Thermo Scientific
Accela RI Detector Hardware Manual
13
5. Holding the detector in place on the benchtop with one hand, use the 5 mm hexagonal
wrench to loosen the two transport screws by two turns.
Do not remove the screws.
6. Move the detector back to a secure position on the benchtop and turn it so that the front
is aligned with the front of the benchtop.
Placement and Environmental Requirements
Place the Accela RI Detector on a benchtop as close as possible to the chromatographic
column outlet. This location minimizes the length of tubing necessary for connection to the
flowcell inlet. If you are using other Accela modules, place the Accela RI Detector on top of
the autosampler as shown in
.
Ensure that the location meets the following requirements:
• A
draft-free
location away from an open window, air conditioner vents, or other
circulating air source.
• A stable room temperature (
±
0.5 °C) for applications requiring maximum detection
sensitivity.
• Clearance of at least 15 cm (6 in.) between the back panel of the Accela RI Detector and
any wall or obstruction. This clear space provides access to the back-panel connectors and
a free flow of cooling air.