Hardware Upgrades
Seal Wash Pump Installation
48
Accela 600 and 1250 Pump Hardware Manual
Thermo Scientific
Seal Wash Pump Installation
If your applications use highly buffered mobile phases, Thermo Fisher Scientific recommends
installing the optional seal wash pump.
To install the optional seal wash pump, you must have the following tools.
To install the optional seal wash pump
1. Turn off the power to the LC pump and unplug the pump’s power cord from line power.
2. Connect the seal wash pump to the pump’s front panel as follows:
a. Using a 2.5 mm hex ball driver, remove the two hex-socket screws that secure the
blind cover to the front panel of the pump (see
). Set the screws aside.
Tool
Use
#2 Phillips head
screwdriver
To disconnect the LED cover plate from the inside of the pump’s
left door and the chassis from the enclosure
2.5 mm hex ball driver
To disconnect the screws that secure the cover plate to the front
panel and secure the seal wash pump to the front panel
CAUTION
To prevent electrical shock, turn off the pump and disconnect it from line
power before you install the seal wash pump.
Note
You use these screws in
to secure the seal wash pump to
the front panel of the pump.