System error list
Code
Message
Context
Description
Recovery action
0x8855
Unknown
initialization error
Startup
Error was detected
during instrument
startup
Export service logs
and send to tech
support for further
assistance
0x8857
Unable to decrypt
upgrade package
Upgrade
Unable to decrypt
the software update
package
Export service logs
and send to tech
support for further
assistance
0x8860
Package is not
suitable for this
instrument
Upgrade
Software update file
is not compatible
with instrument
Export service logs
and send to tech
support for further
assistance
0x8861
Instrument class
is not supported
by this upgrade
package
Upgrade
Software update file
is not compatible
with instrument
Export service logs
and send to tech
support for further
assistance
0x8865
Unable to power up
instrument
Startup
Error was detected
during power up
sequence
• Power cycle
and check if
the instrument
can self
recover
• Export service
logs and send
to tech support
for further
assistance
0x8867
Unable to recover
from upgrade failure
Upgrade
Error was detected
when trying to
recover from an
interrupted software
update
• Ensure that
power is not
turned OFF
during software
update
• Retry software
update,
contact tech
support if error
persists
0x8870
Emergency stop
activated at startup
Startup
Emergency stop
was active at
startup
Reset emergency
stop switch and
perform a power
cycle on the
instrument
Appendix B
Error lists
System error list
B
CTS
™
DynaCellect
™
Magnetic Separation System User Guide
163