Scope of system
CTS
™
Rotea
™
Counterflow Centrifugation System is a stand alone instrument with an operational
configuration comprising:
• The instrument hardware
• Firmware and software
• Dell
™
2
‑
in
‑
1 laptop
• The installed CTS
™
Rotea
™
Single-Use Kit and attached tubing and bags/vessels
• The protocol file
• In-process data that is supplied as protocol metadata before the protocol commences
Instrument hardware configuration
The Rotea
™
instrument's embedded control has been designed to enable integration of functional
components that are external to the base instrument and hence would vary the configuration from the
standard:
• External components may be externally powered
• Communications from the Rotea
™
instrument's embedded control to external components is
through a CAN-BUS interface
• External devices may be individual devices already defined in the Rotea
™
instrument's embedded
schema such as a pump, centrifuge, push button, bubble sensor or valve
• External devices may be newly defined with a set of device specific commands and status reports
• External devices may also comprise a series of components managed by a self-contained control
that responds to purpose designed messages from the Rotea
™
instrument's embedded through the
CAN-BUS
Single-Use Kit
The Single-Use Kit defines:
• The fluidic connections between instrument components such as the pump and valves
• The size of the pump tubing for each pump
• Optionally the volume capacity of each attached bag/vessel
• Optionally the priming volume for each valve line
• Optionally the required minimum volume for each reagent supply to be attached
The Single-Use Kit is identified by:
• A barcode read from a label attached to the kit
• The kit identification includes a unique number, batch identification number and kit type specifier.
The kit type specifier is used for verification of kit type to the protocol requirements.
The scope of a Single-Use Kit may extend beyond the base instrument to include attached devices
such as external mixers and pumps, or simply the tubing or other features that facilitate connection to
external systems such as bio-reactors or electroporation systems that are not managed by the Rotea
™
instrument's control.