5
Aerosol-tight Applications
Basic Principles
5-2
75006476 LH-4000W
Thermo Scientific
Basic Principles
•
Check that the sample containers are well suited for the desired centrifugation process.
Closing aerosol-tight
Rectangular Bucket 75006478
Grease the seals if necessary
,
before closing the bucket. Use lubricant 76003500.
Flap both levers upwards.
The cap can
be
placed easily on the bucket.
Close the bucket aerosol-tight by flapping both levers down until you can hear them locking into
place.
Figure 5-1.
Rectangular Bucket with Cap
CAUTION
Aerosol-tight rotors and tubes may only be opened in an approved
safety workbench when centrifuging dangerous samples.
Mind the maximum permissible load.
CAUTION
Be sure to check all sealings before starting any aerosol-tight applications.
CAUTION
If the levers are not flapped down properly can damage the bucket during
centrifugation.
The bucket is not closed aerosol-tight if the levers are not in their locking position.
Never lift the bucket on the levers.