5. Close the occlusion bed by pushing up on the recessed area at the
bottom, Figure 2-2, until the Locking Tabs snap into place against the
Bed Stops, Figure 2-2, on the housing.
Figure 2-2.
Closing Occlusion Bed
6. Insert the tubing into one of the Tubing Retainers, Figure 2-1, by
moving the retainer toward the center of the pump and placing the
tubing in the v-notch of the retainer. Release the retainer so that the
tubing is secured in place. Repeat for the other Tubing Retainer. The
tubing should be stretched slightly around the rollers so that there is
no excess tubing between the Rollers and the retainers—any excess can
be removed by pulling slightly on the tubing outside of the pump.
CAUTION: Do not close the occlusion bed by pulling up on
the locking tabs. Damage to the tabs may occur.
7. Close the door. The Locking Tabs must be on the outside of the door
and the door fully closed for proper operation.
FH10 MICROFLEX
®
Pump System Operating Manual 2-3
Thermo Scientific
Section 2
Installation and Setup
Installing Tubing
in Pump Head
(continued)
Locking Tab
Locking
Tab
Recess –
Push Here
to Close
Bed
Occlusion
Bed
Bed Stop
Bed Stop