Operation
Instrument Controls Menu
3-44
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
ALLOW OVER/UNDER RANGE:
CURRENTLY: ON
SET TO: OFF ?
TOGGLE VALUE
RANGE AVG DIAGS ALARM
The Analog Output Configuration menu displays a list of the analog
output configuration choices, from which the user selects the parameter
to adjust for the selected output channel. Configuration choices include
selecting range, setting minimum/maximum values, and choosing the
signal to output.
l
In the Main Menu, choose Instrument Controls > I/O
Configuration > Analog Output Config >
All Voltage Channels, All
Current Channels, Voltage Channel 1–6
or
Current Channel 1–6
.
ANALOG OUTPUT CONFIG:
>SELECT RANGE
SET MINIMUM VALUE
SET MAXIMUM VALUE
CHOOSE SIGNAL TO OUTPUT
RANGE AVG DIAGS ALARM
The Select Range screen is used to select the hardware range for the
selected analog output channel. Possible ranges for the voltage outputs
are: 0-100 mV, 0-1, 0-5, and 0-10 V. Possible ranges for the current
outputs are: 0-20 mA and 4-20 mA.
l
In the Main Menu, choose Instrument Controls > I/O
Configuration > Analog Output Config > select Channel >
Select
Range
.
SELECT OUTPUT RANGE:
SELECTED OUTPUT: V ALL
CURRENTLY: 0-10V
SET TO: 0-100mV ?
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
Analog Output Configuration
(Select Action)
Select Range
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......