Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
3-33
The RS-232/RS-485 Selection screen allows the user to choose between
the RS-232 or RS-485 specification for serial communication.
Equipment Damage
Disconnect the serial cable before changing the
RS-232 and RS-485 selection to prevent damage to any equipment
currently connected to the analyzer.
s
l
In the Main Menu, choose Instrument Controls > Communication
Settings > Serial Settings >
RS-232/485 Selection
.
RS-232/RS-485 SELECT:
XXXXXX
** WARNING **
XXXXXXX
DISCONNECT THE SERIAL
X X
CABLES BEFORE CHANGING
X
THE SELECTION!
XXXXXX
TO CONTINUE
RANGE AVG DIAGS ALARM
RS-232/RS-485 SELECT:
CURRENTLY: RS-232
SET TO: RS-485 ?
MAKE SURE THAT THE CABLE
XX
IS OFF: PRESS TO CONFIRM
TOGGLE VALUE
RANGE AVG DIAGS ALARM
The Instrument ID screen allows the operator to edit the instrument
ID number. The ID is used to identify the instrument when using
the C-Link or MODBUS protocols to control the instrument or
collect data. It may be necessary to edit the ID number if two or
more instruments of the same model are connected to one computer.
Valid instrument ID numbers are from 0 to 127. The Model 5028
i
has a default instrument ID of 28. For more information about the
instrument ID, see Appendix B “
l
In the Main Menu, choose Instrument Controls > Communication
Settings >
Instrument ID
.
INSTRUMENT ID:
CURRENTLY: 28
SET TO: 28 ?
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Gesytec Serial Number screen is used to set the serial number for
the Gesytec protocol. The Gesytec serial number defaults to zero. For
more information about the serial number, see Appendix D, “Gesytec
(BH) Protocol”.
l
In the Main Menu, choose Instrument Controls > Communication
Settings >
Gesytec Serial No
.
RS-232/RS-485 Selection
Instrument ID
Gesytec Serial Number
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......