Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
3-19
MASS LIMIT:
CURRENTLY: 1500
SET TO: 150
1
?
MOVE CURSOR
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Next Time screen is used to view and set the initial date and time
(24-hour format) for the next filter tape change.
Note
Channel B settings please refer to Channel A as following
description.
s
l
In the Main Menu, choose Instrument Controls > Filter Tape
Control > Channel A or Channel B>
Next Time
.
START DATE AND TIME:
16 AUG 2016 00:00 ?
SETTING: DAYS
SET MONTHS
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
START DATE AND TIME:
16 AUG 2016 00:00
PRESS TO EDIT
RANGE AVG DIAGS ALARM
The Period screen allows the user to set the period. The next filter tape
change would be scheduled after the number of hours the period is set
for. The official tape change for TUV and U.S. EPA for PM2.5 and
PM10 is set for 8 hours.
Note
Channel B settings please refer to Channel A as following
description.
s
l
In the Main Menu, choose Instrument Controls > Filter Tape
Control > Channel A orChannel B>
Period
.
PERIOD:
CURRENTLY: 24 HRS
SET TO: 00000
8
HRS?
MOVE CURSOR
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
Next Time
Period
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......