Installation
Connecting External Devices
2-12
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
15. Using the available 3/8-inch green vacuum tubing, connect the
intake of the pump to the tee-union fitting at the rear panel of the
5028
i
.
16. Plug the pump power into the PUMP receptacle at the rear of the
5028
i
base instrument. Need a rotation by hand to make the PUMP
connector fixed on PUMP socket.
17. Connect the connector of 3-way valve of CHA and CHB to relative
socket marked with CHA and CHB.
Several components are available for connecting external devices to
iSeries instruments.
These connection options include:
l
Individual terminal board PCB assemblies (standard)
l
Terminal block and cable kits (optional)
l
Individual cables (optional)
For detailed information on the optional connection components, refer
” chapter. For associated part numbers,
External Device Connection Components
.
The terminal board PCB assemblies are circuit boards with a D-Sub
connector on one side and a series of screw terminals on the other side.
These assemblies provide a convenient mechanism for connecting wires
from a data system to the analyzer’s I/O connectors.
The following terminal board PCB assemblies are available for iSeries
instruments:
l
I/O terminal board PCB assembly, 37 pin (standard)
l
D/O terminal board PCB assembly, 37 pin (standard)
l
25-pin terminal board PCB assembly, (included with optional I/O
expansion board)
shows the recommended method for attaching the cable
(customer supplied) to the terminal board using the included tie-down
and spacer.
identifies the connector pins and associated
signals.
Note
Not all of the I/O available in the instrument is brought out on
the supplied terminal board. If more I/O is desired, an alternative means
of connection is required. See optional “
s
Connecting External
Devices
Terminal Board PCB
Assemblies
I/O Terminal Board
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......