MODBUS Protocol
Function Codes
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
C-5
Response
Field Name
(Hex)
Function
0x01
Byte Count
0x03
Output Status 2–10
0xCD
Output Status 11–15
0x0A
The status of outputs 2–10 is shown as the byte value 0xCD, or binary
1100 1101. Output 10 is the MSB of this byte, and output 2 is the
LSB. By convention, bits within a byte are shown with the MSB to the
left and the LSB to the right. Thus, the outputs in the first byte are ’10
through 2’, from left to right. In the last data byte, the status of outputs
15-11 is shown as the byte value 0x0A, or binary 0000 1010. Output
15 is in the fifth bit position from the left, and output 11 is the LSB of
this byte. The four remaining high order bits are zero filled.
Read Holding/Input Registers reads the measurement data from the
instrument. Issuing either of these function codes will generate the same
response. These functions read the contents of one or more contiguous
registers.
These registers are 16 bits each and are organized as shown below. All
of the values are reported as 32-bit IEEE standard 754 floating point
format. This uses 2 sequential registers, least significant 16 bits first.
The request specifies the starting register address and the number of
registers. Registers are addressed starting at zero. Therefore, registers
numbered 1–16 are addressed as 0–15. The register data in the response
message are packed as two bytes per register, with the binary contents
right justified within each byte. For each register, the first byte contains
the high order bits and the second contains the low order bits.
Request
Function Code
1 Byte
0x03 or 0x04
Starting Address
2 Bytes
0x0000 to maximum allowed by instrument
Quantity of Registers
2 Bytes
1 to maximum allowed by instrument
(0x03/0x04) Read Holding
Registers/Read Input
Registers
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......