C-Link Protocol Commands
I/O Configuration
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
B-59
1 pm
3 baro
4 vac
6 pflow
8 ambrt
9 srh
11 ambtemp
12 brdtemp
13 ftemp
15 stemp
17 frol
19 beta
28 braw
29 araw
30 avgpm
32 mass
34 bzero
45 exfg
50 bref
52 cflg
53 aflg
relay stat
This command reports the current relay logic as normally “open” or
normally “closed,” if all the relays are set to same state, that is all open or
all closed. The following example shows that the status of all the relays’
logic is set to normally “open”.
Send:
relay stat
Receive:
relay stat open
Note
If individual relays have been assigned different logic, then the
response would be a 4-digit hexadecimal string with the least significant
byte (LSB) being relay no 1.
s
For example:
Receive:
relay stat 0x0001
(indicates relay no 1 is set to normally open logic, all others are normally
closed)
Receive:
relay stat 0x0005
(indicates relay no 1 and 3 are set to be normally open logic, all others
are normally closed)
set relay open
set relay open
value
set relay closed
set relay closed
value
These commands set the relay logic to normally open or closed for relay
number
value
, where
value
is the relay between 1 and 16. The following
example sets the relay no 1 logic to normally open.
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......