C-Link Protocol Commands
Communications Configuration
B-50
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
Send:
set addr nm 255.255.252.0
Receive:
set addr nm 255.255.252.0 ok
addr ntp
This command reports the IP address for the network time protocol
server. See “Network Time Protocol Server” in the Communications
Settings” section of the “Operation” chapter for more information.
Send:
addr ntp
Receive:
addr ntp 10.209.43.237
set addr ntp
address
This command sets the network time protocol server
address
, where
address
consists of four numbers ranging from 0-255 inclusive, separated
by “.”.
Send:
set addr ntp 10.209.43.237
Receive:
set addr ntp 10.209.43.237 ok
allow mode cmd
This command reports the current allow mode setting: 1 = allow “set
mode local” and “set mode remote” commands; 0 = ignore “set mode
local” or “set mode remote”commands, according to
. The
default value is 0; ignore the commands. The following example shows
that the instrument is configured to ignore “set mode local” or “set
mode remote” commands.
Send:
allow mode cmd
Receive:
allow mode cmd 0
set allow mode cmd
value
This command is used to configure the instrument to
value
, where
value
is either 1 = accept or 0 = ignore the “set mode local” and “set mode
remote” commands, according to
If the instrument is set to accept the commands (
value
= 1), the “set
mode local” command will unlock the instrument and the keypad can
be used to make changes via the front panel.
If the instrument is set to ignore the commands (
value
= 0), the
instrument will respond with “ok” as if the command has been accepted
and acted upon,
but will not change the instrument lock status
(this is
for compatibility with systems expecting and “ok” response).
Note
The instrument will always respond to the command “mode”
with the status of the password lock as “mode local” or “mode remote”,
regardless of the above setting.
s
The following example sets the instrument to accept the “set mode
local” and “set mode remote” commands.
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......