C-Link Protocol Commands
Alarms
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
B-15
alarm conc inst pm min a/b
alarm conc inst pm max a/b
These commands report the current 24-hour instant PM concentration
alarm minimum and maximum values setting. The following example
reports that the instant PM concentration minimum is 100 µg/m
3
.
Send:
alarm conc inst pm min a
Receive:
alarm conc inst pm min a 1.000E+02 µg/m3
Send:
alarm conc inst pm min b
Receive:
alarm conc inst pm min b 1.000E+02 µg/m3
set alarm conc inst pm min a/b
value
set alarm conc inst pm max a/b
value
These commands set the 24-hour instant PM concentration alarm
minimum and maximum values to
value
, where
value
is a floating-
point representation of the concentration alarm limits. Values must be
in the units that are currently set for use. The following example sets the
instant PM concentration alarm maximum value to 10000 µg/m
3
.
Send:
set alarm conc inst pm max a 10000
Receive:
set alarm conc inst pm max a 10000 ok
Send:
set alarm conc inst pm max b 10000
Receive:
set alarm conc inst pm max b 10000 ok
alarm filter tape status a/b
This command reports the status of the filter tape as FAIL or OK. The
following example reports that the filter tape status is OK.
Send:
alarm filter tape status a
Receive:
alarm filter tape status a ok
Send:
alarm filter tape status b
Receive:
alarm filter tape status b ok
alarm flow min a/b
alarm flow max a/b
These commands report the current flow alarm minimum and
maximum value settings. The following example reports that the flow
alarm minimum value is 16.0 LPM.
Send:
alarm flow min a
Receive:
alarm flow min a 16.00 LPM
Send:
alarm flow min b
Receive:
alarm flow min b 16.00 LPM
set alarm flow min a/b
value
set alarm flow max a/b
value
These commands set the flow alarm minimum and maximum values to
value, where value is a floating-point number representing flow alarm
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......