C-Link Protocol Commands
Alarms
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
B-13
representing ambient temperature alarm limits in °C. The following
example sets the ambient temperature alarm maximum value to 38 °C.
Send:
set alarm amb temp max 38.0
Receive:
set alarm amb temp max 38.0 ok
alarm baro pres min
alarm baro pres max
These commands report the current barometric pressure alarm
minimum and maximum value settings. The following example reports
that the barometric pressure alarm minimum value is 500.0 mmHg.
Send:
alarm baro pres min
Receive:
alarm baro pres min 500.0 mmHg
set alarm baro pres min
value
set alarm baro pres max
value
These commands set the barometric pressure alarm minimum and
maximum values to
value
, where
value
is a floating-point number
representing barometric pressure alarm limits in millimeters of mercury.
The following example sets the barometric pressure alarm maximum
value to 800.0 mmHg.
Send:
set alarm baro pres max 800.0
Receive:
set alarm baro pres max 800.0 ok
alarm bench status
This command reports the current status of the bench as FAIL or OK.
The following example reports that the bench status is OK.
Send:
alarm bench status
Receive:
alarm bench status ok
alarm beta count min
alarm beta count max
These commands report the current beta count alarm minimum and
maximum value settings. The following example reports that the beta
count alarm minimum value is 5000.
Send:
alarm beta count min
Receive:
alarm beta count min 5000
set alarm beta count min
value
set alarm beta count max
value
These commands set the beta count alarm minimum and maximum
value
, where
value
is an integer number representing beta count limits.
The following example sets the beta count alarm maximum value to
20000.
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......