Servicing
Radius Tube Replacement
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
7-31
cover.
2. Disconnect cables from the measurement interface board and
motherboard.
3. Unscrew the four mounting screws and lift out the transformer
assembly.
4. Install the new transformer by following the previous steps in
reverse.
Figure 7–18.
Replacing the Transformer
Use the following procedure to replace the radius tube (
).
Equipment Required:
Radius tube assembly
Hex driver, 9/16-inch
Hex driver, 7/64-inch
Equipment Damage
Some internal components can be damaged by
small amounts of static electricity. A properly grounded antistatic wrist
strap must be worn while handling any internal component.
s
1. Turn the instrument OFF, unplug the power cord, and remove
the cover. Channel A and Channel B have the same operation for
replacement.
2. Loosen the two screws of the radius mounting block using a 9/16-
inch hex driver.
3. Loosen the two mounting screws on the measurement head
assembly. Remove tube by slightly turning side-to-side, while
pulling.
Radius Tube
Replacement
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......