Operation
Service Menu
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
3-83
ANALOG INPUT CAL:
>INPUT CHANNEL 1
INPUT CHANNEL 2
INPUT CHANNEL 3
INPUT CHANNEL 4
INPUT CHANNEL 5
RANGE AVG DIAGS ALARM
ANALOG INPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE AVG DIAGS ALARM
INPUT CHANNEL 6
INPUT CHANNEL 7
The Analog Input Calibrate Zero screen allows the user to calibrate the
zero state of the selected analog input.
l
In the Main Menu, choose Service > Analog Input Cal > select
Channel >
Calibrate Zero
.
ANALOG INPUT CAL: ZERO
DISCONNECT SELECTED INPUT!
SELECTED INPUT: INPUT 1
CURRENTLY: 0.04 V
CALIBRATE INPUT TO ZERO
RANGE AVG DIAGS ALARM
The Analog Input Calibrate Full-Scale screen allows the user to calibrate
the full-scale state of the selected analog input.
l
In the Main Menu, choose Service > Analog Input Cal > select
Channel >
Calibrate Full Scale
. (Hook up a voltage source of 10 V
to the analog input channel.)
ANALOG INPUT CAL: SPAN
PROVIDE VOLTAGE TO INPUT!
SELECTED INPUT: INPUT 1
CURRENTLY: 9.84 V
SET TO: 10.0
0
V
CALIBRATE TO VALUE
RANGE AVG DIAGS ALARM
The Language menu allows the user to choose languages include
Chinese and English.
l
In the Main Menu, choose
Service>Language
.
Analog Input Calibrate Zero
Analog Input Calibrate Full-Scale
Language
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......