Operation
Service Menu
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
3-75
The Restore Default Calibration screen allows the user to reset the
default values.
Note
Be sure to close the toggle switch after calibration is complete.
s
l
In the Main Menu, choose Service > Pres/Vacuum Calibration >
Vac/Flow or Vac/Flow B >
Set Defaults
.
RESTORE DEFAULT CAL:
RESTORE
RANGE AVG DIAGS ALARM
RESTORE DEFAULT CAL:
RESTORE
ARE YOU SURE YOU WANT TO?
x
PRESS TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM
The Flow Calibration submenu is used to view and set the flow
calibration flow point. The flow calibration screen is visible only when
the instrument is in service mode. For more information on the service
Note
Flow Span B settings please refer to Flow Span as following
description.
s
Note
This adjustment should only be performed by an instrument
service technician.
s
l
In the Main Menu, choose Service >
Flow Calibration>Flow Span
or Flow Span B
.
FLOW CALIBRATION:
>AUTO
MANUAL
RANGE AVG DIAGS ALARM
The Auto Flow Calibration screen allows the user to view and set the
correct flow rate. The span value is automatically calculated.
l
In the Main Menu, choose Service > Flow Calibration > Flow Span
or Flow Span
B>Auto
.
Restore Default Calibration
Flow Calibration
Auto Flow Calibration
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......