Operation
Service Menu
3-72
Model 5028
i
Instruction Manual
Thermo Fisher Scientific
The Ambient Relative Humidity screen allows the user to view and
calibrate relative humidity. Adjust the zero offset until the RH agrees
with your treaceable standard.
l
In the Main Menu, choose Service > RH/Temp Calibration >
Ambient RH
.
CALIBRATE AMBIENT RH:
REL HUMIDITY: 40.0 %
OFFSET: 0.0 %
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Pressure/Vacuum Calibration submenu allows the user to view and
calibrate the pressure sensor and vacuum flow sensor. The pressure/
vacuum calibration submenu is visible only when the instrument is in
service mode. For more information on the service mode, see “
Note
This adjustment should only be performed by an instrument
service technician.
s
l
In the Main Menu, choose Service >
Pres/Vacuum Calibration
.
PRES/VACUUM CALIBRATION:
>BARO PRES
VAC/FLOW
RANGE AVG DIAGS ALARM
VAC/FLOW B
The Barometer Pressure Calibration submenu is used to calibrate the
barometer pressure span or reset default values.
l
In the Main Menu, choose Service > Pres/Vacuum Calibration >
Baro Pres
.
BARO PRES CALIBRATION:
>OFFSET 0.0
RANGE AVG DIAGS ALARM
SET DEFAULTS
SPAN 1.0000
Ambient Relative Humidity
Pressure/Vacuum
Calibration
Barometer Pressure Calibration
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......