Operation
Alarms Menu
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
3-67
The Flow screen displays the current flow reading and sets the minimum
and maximum alarm limits. Acceptable alarm limits range from -5
mmHg to 40 mmHg. If the flow reading goes beyond the minimum or
maximum limit, an alarm is activated and the alarm (bell) icon appears
in the status bar on the Run screen and in the Main Menu
Note
Flow B settings please refer to Flow as following description.
s
l
In the Main Menu, choose Alarms > Pressure/Vacuum Alarms >
Flow or Flow B
.
FLOW:
ACTUAL 23.1 mmHg
>MIN -5.0 mmHg
MAX 40.0 mmHg
RANGE AVG DIAGS ALARM
The Minimum Flow alarm limit screen is used to change the minimum
flow alarm limit. The minimum and maximum flow screens function
the same way.
l
In the Main Menu, choose Alarms > Pressure/Vacuum Alarms >
Flow or Flow B >
Min
or
Max
.
FLOW PRESSURE:
ACT MIN: -5.0 mmHg
SET MIN: -4.9 mmHg?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Flow Alarms submenu allows the user to view and set the flow
alarm parameters.
l
In the Main Menu, choose Alarms >
Flow Alarms
.
FLOW ALARMS:
>FLOW OK
FLOW B OK
RANGE AVG DIAGS ALARM
Flow
Min and Max Flow Limits
Flow Alarms
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......