Operation
Alarms Menu
Thermo Fisher Scientific
Model 5028
i
Instruction Manual
3-63
l
In the Main Menu, choose Alarms > RH/Temperature Alarms >
Sample RH or Sample RH B >
Min
or
Max
.
SAMPLE RH:
ACT MIN: 5.0 %
SET MIN: 35 %?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Ambient Temperature screen displays the current ambient
temperature value and sets the minimum and maximum alarm limits.
Acceptable alarm limits range from -30 °C to 60 °C. If the ambient
temperature reading goes beyond the minimum or maximum limit, an
alarm is activated and the alarm (bell) icon appears in the status bar on
the Run screen and in the Main Menu.
l
In the Main Menu, choose Alarms > RH/Termperature Alarms >
Ambient Temp
.
AMBIENT TEMP:
ACTUAL -30.0
o
C
>MIN -30.0
o
C
MAX 60.0
o
C
RANGE AVG DIAGS ALARM
The Minimum Ambient Temperature alarm limit screen is used to
change the minimum ambient temperature alarm limit. The minimum
and maximum ambient temperature screens function the same way.
l
In the Main Menu, choose Alarms > RH/Temperature Alarms >
Ambient Temp >
Min
or
Max
.
AMBIENT TEMPERATURE:
ACT MIN: -30.0
o
C
SET MIN: -20.0
o
C?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Flow Temperature screen displays the current ambient temperature
value and sets the minimum and maximum alarm limits. Acceptable
alarm limits range from 5 to 60 °C. If the flow temperature reading goes
beyond the minimum or maximum limit, an alarm is activated and the
alarm (bell) icon appears in the status bar on the Run screen and in the
Main Menu
Ambient Temperature
Min and Max Ambient
Temperature Limits
Flow Temperature
Содержание 5028i
Страница 1: ...Model 5028i Instruction Manual Continuous Particulate Monitor CIC0001582 April 2018...
Страница 2: ......
Страница 3: ...Model 5028i Instruction Manual Continuous Particulate Monitor Part Number CIC0001582 8Apr2018...
Страница 6: ...WEEE Compliance Thermo Fisher Scientific...
Страница 12: ...About This Manual WEEE Symbol vi Model 5028i Instruction Manual Thermo Fisher Scientific...
Страница 36: ......
Страница 142: ......
Страница 154: ......
Страница 220: ......
Страница 222: ...System Description Hardware 8 2 Model 5028i Instruction Manual Thermo Fisher Scientific Figure 8 1 Hardware Components...
Страница 232: ......
Страница 328: ......
Страница 329: ......