C-Link Protocol Commands
I/O Configuration
Thermo Fisher Scientific
Model 49
i
Instruction Manual
B-41
dig in
This command reports the status of the digital inputs as a 4-digit
hexadecimal string with the most significant bit (MSB) being input 16.
Send:
dig in
Receive:
dig in 0xff7f
din
channel
This command reports the action assigned to input
channel
and the
corresponding active state. The example that follows reports the input 5 to
be assigned an index number 9 corresponding to action of “analog outputs
to zero” with the active state being high.
Send:
din 5
Receive:
din 5 9 AOUTS TO ZERO high
set din
channel index state
This command assigns digital input
channel
(1-16) to activate the action
indicated by
index
(1-35), when the input transitions to the designated
state
(high or low). Use “list din var” command to obtain the list of supported
index
values and corresponding actions.
Send:
set din 1 3 high
Receive:
set din 1 3 high ok
dout
channel
This command reports the index number and output variable and the active
state assigned to output
channel.
The example that follows reports the input
4 to be assigned an index number 11 corresponding to “general alarm” with
the active state being open.
Send:
dout 4
Receive:
dout 4 11 GEN ALARM open
set dout
channel index state
This command assigns digital output
channel
to be assigned to the action
associated with
index
, and assigns it an active state of
state
(open or closed).
Send:
set dout 4 11 open
Receive:
set dout 4 11 open ok
dtoa
channel
This reports the outputs of the 6 or 12 digital to analog converters,
according to
. The example that follows shows that the DAC 1 is
97.7% full-scale.
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...