Preventive Maintenance
Ozone Scrubber Test
5-10
Model 49
i
Instruction Manual
Thermo Fisher Scientific
9. Plug the solenoid power line into the appropriate connector on the
measurement interface board. Make sure the solenoid is activated by
choosing
Pressure
from the Diagnostics menu.
10. Note the pressure as P
NC
the solenoid is faulty
11. The solenoid is faulty if either P
NO
or P
NC
is greater than the pressure
determined in the “External Leaks” section described previously.
Ozone Scrubber Test
Use the following procedure to determine the efficiency of the ozone
scrubber. To replace the scrubber, refer to the “Ozone Scrubber
Replacement” procedure in the “Servicing” chapter.
1. Generate a source of ozone of about 0.5 ppm and feed into the
instrument. Note the concentration as C.
2. Press
to display the Main Menu.
3. Press
to scroll to Service > press
to scroll to
Intensity Check > press
to choose Int A Reference Gas.
4. When the frequency stabilizes, note the frequency as FREQ 1.
5. Turn ozonator off and when the frequency stabilizes, note the frequency
as FREQ 2.
6. Determine pressure and temperature, note as P and T.
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...