Preventive Maintenance
Capillary Service
Thermo Fisher Scientific
Model 49
i
Instruction Manual
5-5
1. Press
to display the Main Menu.
2. Use
to scroll to Service, press
>
to scroll to
Intensity Check
and press
.
The Intensity Check screen appears.
The noise value displayed after 20 seconds should be below 4.0 Hz for a
fully warmed-up lamp (see the “Troubleshooting” chapter if the noise is
excessive).
Capillary Service
The capillaries normally only require inspection when performance
indicates there may be a flow problem. Use the following procedure to
service the capillaries (
Equipment Required:
Capillary
Wire, less than 0.015-inch OD
1. Turn instrument OFF, unplug the power cord, and remove the cover.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...