Preventive Maintenance
Lamp Replacement
Thermo Fisher Scientific
Model 49
i
Instruction Manual
5-3
1. Turn off the power and disconnect the power line.
2. Loosen the knurled nut around the tube and carefully slide out tube.
3. Push a piece of lens paper down the tube using a 1/4-inch piece of
Teflon® tubing so as not to damage the tube. Use a cotton swab to clean
the window surfaces through the holes that the tube fits into.
4. Both absorption tubes are identical, so they can be replaced in either
position. Replacement of absorption cells is opposite to that of removal.
Since the Model 49
i
is a ratio instrument, and cleaning the absorption
tubes does not affect the calibration, it is not necessary to recalibrate the
instrument every time the cells are cleaned.
5. Re-install the instrument cover.
If windows are severely contaminated, they are best cleaned by removing the
windows from the bench. The windows on the detector side can be removed
by removing the detector block and carefully removing the windows. The
windows on the source side can be removed by removing the source block to
gain access to the windows. It is recommended that the Model 49
i
be
recalibrated if the optical bench has been completely disassembled. Always
leak-check the system after any component removal.
Lamp Replacement
The lamp control system of the Model 49
i
has been designed to operate the
lamp conservatively to increase its life. However, the lamp should be
replaced when any one of the following conditions occur:
•
No light output.
•
Inability to adjust lamp position to obtain an output detection
frequency of 65 kHz.
•
Noisy output signal, which has been traced to an unstable lamp (see the
“Troubleshooting” chapter).
It is not necessary to recalibrate the Model 49
i
since it is a ratio instrument
and replacing the lamp does not affect the calibration.
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...