Operation
Settings
Thermo Scientific
43iQ Instruction Manual
3-71
The Advanced Measurement Settings screen allows the user to calibrate the
optical bench and set other advanced settings.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings
The Advanced Measurements Settings screen contains the following
information:
●
Optical Bench Settings:
Calibrates detector gain, PMT supply, flash
voltage, and initial flash reference.
●
Flash Lamp:
Enables/Disables flash lamp.
●
Optical Span Test:
Enables/Disables test LED and reads the LED SO
2
concentration.
●
Extended Ranges:
Enables/Disables extended ranges feature. When
changing to extended range from standard range, the PMP voltage
must be readjusted. For more information on readjusting the PMT
voltage, go to Settings>Measurement Settings>Advanced Measurement
Settings>Optical Bench Settings>
Continue to PMT Settings
.
●
Compensation:
Allows the user to compensate for changes in
temperature and pressure concentration.
●
Perm Oven Selection:
Allows the user to select the operating temperature
of the perm oven.
●
Pressure Calibration:
Calibrates pressure.
Advanced Measurement
Settings
Содержание 43iQ
Страница 1: ...43iQ Instruction Manual Sulfur Dioxide Analyzer 117568 00 1Sep2021 ...
Страница 2: ......
Страница 8: ......
Страница 18: ......
Страница 34: ...Operation Instrument Display 3 8 43iQ Instruction Manual Thermo Scientific ...
Страница 162: ......
Страница 166: ...Maintenance Pump Rebuilding 5 4 43iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 170: ......
Страница 180: ......
Страница 188: ...Servicing Fan Replacement 7 8 43iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 228: ......
Страница 238: ......
Страница 262: ......
Страница 270: ......
Страница 282: ......
Страница 283: ......