The Preventive Maintenance screen reminds the user when certain
instrument components need to be serviced/replaced. When the “Months
Left” has decreased to 2, the row is highlighted yellow. If the “Months
Left” is 1 or less, the row is highlighted red and the maintenance icon
(gears) will appear in the status bar located at the bottom of the screen.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Settings>Health Check>Maintenance>Preventive
Maintenance
The Preventive Maintenance screen contains the following information:
●
Across:
●
Component:
Device to be routinely serviced or replaced.
●
Interval in Months:
Expected period of time before a component
needs to be checked and/or serviced.
●
Months Left:
Count down timer in months. Remaining time since
the beginning of the service interval. When the value is 1 or less,
the row will be highlighted and it is suggested that the component
should be checked and/or serviced.
●
Reset:
Once the component is serviced/replaced, the user presses the
Reset button and the “Months Left” value resets to the “Service
Interval in Months” value.
●
Alert:
Allows the user to opt out of receiving preventive
maintenance alerts. Displays Enabled/Disabled for each
component.
Preventive Maintenance
Содержание 43iQ
Страница 1: ...43iQ Instruction Manual Sulfur Dioxide Analyzer 117568 00 1Sep2021 ...
Страница 2: ......
Страница 8: ......
Страница 18: ......
Страница 34: ...Operation Instrument Display 3 8 43iQ Instruction Manual Thermo Scientific ...
Страница 162: ......
Страница 166: ...Maintenance Pump Rebuilding 5 4 43iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 170: ......
Страница 180: ......
Страница 188: ...Servicing Fan Replacement 7 8 43iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 228: ......
Страница 238: ......
Страница 262: ......
Страница 270: ......
Страница 282: ......
Страница 283: ......