Servicing
Optional Manifold Replacement
7-40
43iQ Instruction Manual
Thermo Scientific
4.
Replace the kicker assembly and assemble in reverse order.
Use the following procedure to replace the manifold.
Equipment required:
Hex wrench, 9/16
Hex drive, 9/64
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover (
).
2.
Unplug three electrical connections (J5, J6, and J8) from the step pol
board 1.
Figure 7–45.
Replacing the Manifold pt 1
3.
Unfasten three nuts. Remove the nuts, front and back ferrules as shown
from span, zero in, sample back panel (
).
Optional
Manifold
Replacement
Содержание 43iQ
Страница 1: ...43iQ Instruction Manual Sulfur Dioxide Analyzer 117568 00 1Sep2021 ...
Страница 2: ......
Страница 8: ......
Страница 18: ......
Страница 34: ...Operation Instrument Display 3 8 43iQ Instruction Manual Thermo Scientific ...
Страница 162: ......
Страница 166: ...Maintenance Pump Rebuilding 5 4 43iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 170: ......
Страница 180: ......
Страница 188: ...Servicing Fan Replacement 7 8 43iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 228: ......
Страница 238: ......
Страница 262: ......
Страница 270: ......
Страница 282: ......
Страница 283: ......