Servicing
DMC Optical Bench Removal
Thermo Scientific
43iQ Instruction Manual
7-37
Use the following procedure to remove and replace the photo lamp
detector board.
Equipment required:
Phillips drive, #1 and #2
1.
Unplug cable.
2.
Using #2 Phillips drive, unfasten three #6-32 pan head screws.
Figure 7–42.
Replacing the Photo Lamp Detector Board pt 1
3.
Pull detector cover outwards.
4.
Using #1 Phillips drive, unfasten four #4-40 pan head screws.
5.
Pull photo lamp detector board outwards.
6.
Replace lamp detector board and assemble in reverse order.
Photo Lamp
Detector Board
Replacement