Operation
Calibration
3-14
42iQD Instruction Manual
Thermo Scientific
The Calibrate NO
x
Span Coefficient screen is used to enter span
concentration and calibrate the NO
x
span coefficient. The span coefficient
is calculated, stored, and used to correct the current reading.
The following screens are shown in single range mode and dual or auto
range mode. In dual or auto range modes, “High” or “Low” is displayed to
indicate the calibration of the high or low coefficient. The Calibrate High
Span Coefficient and Calibrate Low Span Coefficient screens function the
same way.
It is important to note the averaging time when calibrating. The longer the
averaging time the more precise the calibration results. To achieve
maximum precision, allow the instrument to stabilize each time input gas is
changed and set the averaging time to 300-second averaging.
Home Screen>Calibration>Calibrate Span Coefficient (single range
mode)
Home Screen>Calibration>Calibrate High Range Coefficient (dual or
auto range mode)
Calibrate NO
x
Span
Coefficient
Содержание 42iQD
Страница 1: ...42iQD Instruction Manual NOx Analyzer 119013 00 1Sep2021 ...
Страница 2: ......
Страница 32: ...Operation Instrument Display 3 8 42iQD Instruction Manual Thermo Scientific ...
Страница 182: ...Maintenance Pump Rebuilding 5 8 42iQD Instruction Manual Thermo Scientific Figure 5 3 Pump Rebuilding ...
Страница 184: ......
Страница 198: ...Servicing Fan Replacement 7 8 42iQD Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 212: ...Servicing Power Supply Replacement 7 22 42iQD Instruction Manual Thermo Scientific Figure 7 26 Replacing Power Supply ...
Страница 235: ...Servicing DMC Ozonator Thermo Scientific 42iQD Instruction Manual 7 45 Figure 7 49 Flow Switch Remove ...
Страница 238: ......
Страница 272: ......
Страница 284: ......
Страница 285: ......