Servicing
DMC Ozonator
7-40
42iQD Instruction Manual
Thermo Scientific
Use the following procedure to replace the ozonator and transformer board,
and to replace the flow switch.
●
Removing the ozonator and transformer
●
Replacing the ozonator and transformer board
●
Replacing the flow switch
Use the following procedure to remove the ozonator and transformer.
Equipment Required:
Phillips drive, #2
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover (
).
2.
Unplug converter DMC (J2) and flow pressure board (J4).
Figure 7–44.
HVPS, Electrical
DMC Ozonator
Ozonator and
Transformer
Removal