Level Monitor

4-1

Thermo 

Scientific

Section

 4  

Installation

Factory Installation

The level monitor is not normally mounted at the factory when purchased
with a storage tank. Connect the level monitor electrically as shown on the
wiring diagram included with the equipment. Adjust the level monitor as
described in the ADJUSTMENT section of this manual. Refer to the
TECHNICAL CHARACTERISTICS section for the switch contact full load
rating limits for the external equipment.

Field Installation

The following procedures are the recommended method(s) for installing the
level monitor unit(s) when not factory installed.

Note

To ensure proper installation, installation should be done by qualified

personnel. Read these procedures thoroughly. If any portion of them is unclear
to you or you have unanswered questions concerning them, contact Technical
Services before proceeding. 

1.   Install a 3/8 inch NPT tin-lined (for distilled water storage tanks) spud 24

inches (610 mm) from the top of the tank as shown in Figure 4-1. If the
spud cannot be located 24 inches (610 mm) from the top of the tank,
locate the spud as close to 24 inches (610 mm) as physically possible.

Note 

Installation in a Thermo Scientific tank (cylindrical), refer to the tank

manual to properly install the Level Monitor in the proper port. 

2.   Install the U-tube assembly in the spud; then install the level monitor

controller into the U-tube assembly using Teflon® ribbon tape on the
piping connections.

C

aution 

Do not tighten the level monitor by hand by holding the case. To

avoid leaks and damage to the case, use a parallel jaw wrench on the level
monitor’s hexagonal fitting. 

3.   Connect the level monitor to the LWCO box (twist lock cord) and the

power side to a 12W receptacle per the Wiring Diagram provided in the
enclosed operating manual on CD. Refer to the TECHNICAL
CHARACTERISTICS section for the switch contact full load rating limits
for the external equipment.

4.   Adjust the level monitor as described in the ADJUSTMENT and

MAINTENANCE sections in the operating manual.

Note 

Level Monitor has been preset to the following set points. SW1 = 17”

and DB1 = 7”. Refer to Programming Flow Chart to make adjustments to suit
your tank. 

Содержание 2112 Series

Страница 1: ...Installation Guide 7991048 Rev 1 For further information refer to the enclosed operating manual on CD A printed copy of the operating manual is available from Technical Services Visit us online to reg...

Страница 2: ...entific MANUAL NUMBER 7991048 1 41865 30 03 2018 Programming instructions flow chart updated Figs updated with new controller TFI 0 01 04 2013 Initial release CCS REV ECR ECN DATE DESCRIPTION BY Level...

Страница 3: ...this symbol Equipment being maintained or serviced must be turned off and locked off to prevent possible injury Hot surface s present which may cause burns to unprotected skin or to materials which ma...

Страница 4: ...Table of Contents Introduction and Safety Information 1 1 Programming 2 1 Technical Characteristics 3 1 Installation 4 1 Level Monitor ii Thermo Scientific Section 1 Section 2 Section 3 Section 4...

Страница 5: ...ount the level monitor on the storage tank Functionally the Level Monitor operates as a pressure controlled electrical switch which changes state on rising water head pressure This manual contains imp...

Страница 6: ...P CNT FILTER NO TRIP DELAY SAVE CHNG YES SAVING PROGRAMMING FLOW CHART DIRECT ACTING LEVEL t y t y Press and Hold Simultaneously X4 WC ADJ SP DB IAW X XX PASSWORD 0000 X4 OK CLR MAX MIN SET UNITS WC S...

Страница 7: ...s the button 9 The display shows SET UNITS then press the button 10 The display shows SW1 then press the button 11 If the display shows OPEN RISE then press the button if not then press the button unt...

Страница 8: ...then press the button 27 The display shows FILTER then press the button 28 The display shows TRIP DELAY then press the button 29 The display shows SAVE CHNG then press the button 30 The display shows...

Страница 9: ...FLOW CHART REVERSE ACTING LEVEL MONITOR Press and Hold Simultaneously X4 WC ADJ SP DB IAW X XX PASSWORD 0000 X4 OK CLR MAX MIN SET UNITS WC SET UNITS SW1 CLOSE RISE Level Monitor 2 4 Thermo Scientifi...

Страница 10: ...n press the button 10 The display shows SW1 then press the button 11 If the display shows OPEN RISE then press the button if not then press the button until the display shows CLOSE RISE 12 Then press...

Страница 11: ...button 26 The display shows RESET TRIP CNT then press the button 27 The display shows FILTER then press the button 28 The display shows TRIP DELAY then press the button 29 The display shows SAVE CHNG...

Страница 12: ...conduit Weight 4 lb 12oz 2 15 kg Electrical Rating 90 130 VAC as power source TB1 Terminal Block accepts 14 22 AWG wiring Switch Contact Ratings in amperes at 50 60 Hz Maximum load rating is 10 Amps 7...

Страница 13: ...shown in Figure 4 1 If the spud cannot be located 24 inches 610 mm from the top of the tank locate the spud as close to 24 inches 610 mm as physically possible Note Installation in a Thermo Scientifi...

Страница 14: ...4 Installation Figure 4 1 Direct Acting Level Monitor Installation used in operation with Stills and FAC Controls Figure 4 2 Reverse Acting Level Monitor Installation used in distribution pump assembl...

Страница 15: ...er Scientific and its subsidiaries Specifications terms and pricing are subject to change Not all products are available in all countries Please consult your local sales representative for details The...

Отзывы: