Servicing
Pressure Regulator(s) Adjustment
Thermo Fisher Scientific
Model 146
i
Instruction Manual
7-51
11. Remove defective flow controller from the bracket by removing the two
retaining screws.
12. Install the new flow controller by following the previous steps in reverse.
13. Check the system for leaks. Refer to “Leak Test and Pump Checkout” in
this chapter.
14. Recalibrate.
Pressure Regulator(s)
Adjustment
Use the following procedure to adjust the pressure regulator and the back
pressure regulator.
Equipment required:
Digital flow meter capable of measuring 150 sccm
1. Configure a digital flow meter (capable of measuring 150 sccm) with a
0.01-inch capillary restricting the meter input.
2. For instruments equipped with the ozonator option, attach the digital
flow meter to the gas line immediately preceding the ozonator capillary
holder (attached to the solenoid valve on the ozonator).
For instruments equipped with only the permeation oven option, attach
the digital flow meter in place of the tee fitting between the back
pressure regulator input and the Perm 1 solenoid valve.
WARNING
The service procedures in this manual are restricted to qualified
service representatives.
If the equipment is operated in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...