Servicing
Photometer Lamp Voltage Adjustment (Photometer option)
Thermo Fisher Scientific
Model 146
i
Instruction Manual
7-33
Photometer Lamp
Voltage Adjustment
(Photometer option)
Use the following procedure to adjust the photometer lamp intensity to
approximately 100,000 Hz.
Note
After turning power ON, wait for the lamp to stabilize (approximately
15 minutes) before beginning the following lamp voltage adjustment
procedure.
1. From the Main Menu, press
to scroll to Service > press
>
to scroll to
Lamp Setting
> and press
.
The Photometer Lamp Setting screen appears.
Note
If Service is not displayed, refer to “
, then return to the beginning of this step.
2. At the Photometer Lamp Setting screen, use
to
increment/decrement the Lamp Setting % until Intensity indicates
approximately 100,000 Hz.
3. Press
to save the setting.
Photometer Lamp
Heater Replacement
Use the following procedure to replace the photometer lamp heater
(
).
Equipment required:
Photometer lamp heater
Allen wrench, 7/64-inch
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Disconnect the plumbing and electrical connections.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...