Start-Up
Unlock
Counterweight
The window counterweight is secured to the back of the cabinet to protect
it from damage during shipping.
Caution
The counterweight on
the cabinet back compensates for
the weight of the front window.
Do not try to move the front
window before the counterweight
has been unlocked.
s
Warning
After unlocking, keep
hands and fingers away from the
motion range of the
counterweight!
s
To unlock the counterweight,
remove the shipping screws
(see location arrows in Figure 3-2)
and retain them for future
shipping needs. A Torx wrench is
shipped in the manual bag for
this purpose.
Warning
One person should hold the counterweight while a second
person removes the screws. This facilitates screw removal and avoids the
possibility of stripping the screws.
s
Note
The cabinet should be leveled only after it has been located in the
desired working location.
s
For cabinets without a stand, place a bubble level onto the work tray
surface and adjust the cabinet until the level indicates an absolutely
horizontal position in all directions.
For cabinets with an optional stand, place a bubble level onto the work
tray and adjust the four levelers of the stand using an open end or
adjustable wrench until the bubble level indicates an absolutely horizontal
position in all directions. Check from right to left, and from back to front.
Level the Cabinet
Counterweight
Figure 3-2.
Unlocking the Counterweight
Содержание 1300 Series
Страница 25: ...Thermo Scientific Section 2 Installation 2 8 1300 Series A2 ...
Страница 31: ...Thermo Scientific Section 3 Start Up 3 6 1300 Series A2 ...
Страница 39: ...Thermo Scientific Section 4 Operation 4 8 1300 Series A2 ...
Страница 45: ...Thermo Scientific Section 5 Cleaning and Decontamination 5 6 1300 Series A2 ...
Страница 69: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...
Страница 70: ...1300 系列 A2 II级 A2型 生物安全柜 操作手册 5991355 ...
Страница 91: ...Section 2 安装 维修阀安装 续 维修阀 维修阀插头 垫片 螺母 图2 4 阀门总成和安装 2 6 1300 Series A2 Thermo Scientific 螺纹管 底座 ...
Страница 129: ...仪器类型 部件号 序列号 维修号 地点 操作者标注 技术员 公司 记录 日期 签名 Thermo Scientific 1300 Series A2 9 1 ...