The test conditions according to YY 0569-2005 were performed with the
standard version of the safety cabinet, without exhaust system accessories.
An available accessory for the biological safety cabinet exhaust system is an
exhaust thimble connection for the connection to an on-site exhaust
system with blower.
The exhaust system components are installed to the exhaust opening on
top of the cabinet. The threaded inserts for the retaining screws are
pre-installed in the unit. The gap on the exhaust transition should be
nominally set at 50 mm (2 in.).
Installation on a 1.2 m model:
1. Turn the unit off and disconnect it from the power source.
2. Thoroughly clean the surfaces around the exhaust opening and of the
accessory to be installed, to
ensure that they are free of debris.
3. Apply sealant to the contact
surfaces.
4. Secure the accessory by tightening
the supplied retaining screws
(M5) wrench-tight.
5. Using the adjustment nuts on the
sliding panel, set the gap height
to an opening of 2 inches.
6. Remove any excess sealant.
2-2
1300 Series A2
Thermo Scientific
Section 2
Installation
Optional Exhaust
Transition
Exhaust System
Accessories
Thimble
Retaining
screws
Exhaust
opening
Threaded
inserts
Figure 2-1.
1.2 m Thimble
Содержание 1300 Series
Страница 25: ...Thermo Scientific Section 2 Installation 2 8 1300 Series A2 ...
Страница 31: ...Thermo Scientific Section 3 Start Up 3 6 1300 Series A2 ...
Страница 39: ...Thermo Scientific Section 4 Operation 4 8 1300 Series A2 ...
Страница 45: ...Thermo Scientific Section 5 Cleaning and Decontamination 5 6 1300 Series A2 ...
Страница 69: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...
Страница 70: ...1300 系列 A2 II级 A2型 生物安全柜 操作手册 5991355 ...
Страница 91: ...Section 2 安装 维修阀安装 续 维修阀 维修阀插头 垫片 螺母 图2 4 阀门总成和安装 2 6 1300 Series A2 Thermo Scientific 螺纹管 底座 ...
Страница 129: ...仪器类型 部件号 序列号 维修号 地点 操作者标注 技术员 公司 记录 日期 签名 Thermo Scientific 1300 Series A2 9 1 ...