The sample chamber is accessible through various positions of the front
window. The working position of the window is a 250 mm opening. For
loading the chamber, the window opens fully. For decontamination and
energy saving, enter into Night-Set-Back mode by closing the window.
When the window is closed, the blower speed is reduced, minimizing
energy consumption and noise emittance without compromising
particulate containment inside the work chamber.
Note
The working and loading window positions are indicated on the user
interface, as controlled by micro switches and displayed by status
indicators on the control panel.
s
Use of the Window
1-6
1300 Series A2
Thermo Scientific
Section 1
Description
Window
Window
Pillar dress
panel
Opening height
for cleaning window
Pillar dress
panel
Closed position
Work
position
Max opening for
loading samples
1300 SERIES A2
1300 SERIES A2
Figure 1-4.
Window Openings
Содержание 1300 Series
Страница 25: ...Thermo Scientific Section 2 Installation 2 8 1300 Series A2 ...
Страница 31: ...Thermo Scientific Section 3 Start Up 3 6 1300 Series A2 ...
Страница 39: ...Thermo Scientific Section 4 Operation 4 8 1300 Series A2 ...
Страница 45: ...Thermo Scientific Section 5 Cleaning and Decontamination 5 6 1300 Series A2 ...
Страница 69: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...
Страница 70: ...1300 系列 A2 II级 A2型 生物安全柜 操作手册 5991355 ...
Страница 91: ...Section 2 安装 维修阀安装 续 维修阀 维修阀插头 垫片 螺母 图2 4 阀门总成和安装 2 6 1300 Series A2 Thermo Scientific 螺纹管 底座 ...
Страница 129: ...仪器类型 部件号 序列号 维修号 地点 操作者标注 技术员 公司 记录 日期 签名 Thermo Scientific 1300 Series A2 9 1 ...