4 • Troubleshooting
Doc. 065378-01 1/12
85
•
Incorrect reference wavelength selected
Select a different reference wavelength. For guidelines to follow when
selecting the reference wavelength, see
.
•
Bandwidth too wide
Select a bandwidth that provides the resolution required to achieve the desired
sensitivity. Narrow spectra peaks require narrow bandwidths. A wider
bandwidth reduces baseline noise, but may reduce peak height.
•
Reference bandwidth too wide
Select a bandwidth that provides the resolution required to achieve the desired
sensitivity. A wider bandwidth reduces baseline noise, but may increase the
chance that the sample absorbs in the reference bandwidth, thus causing
reduced sensitivity and peak height.
•
Cell flow is reversed
Check that the flow cell inlet tubing is routed through the heat exchanger (see
) and is correctly connected to the inside front panel (see
). Check that solvent flow is directed to the cell inlet and out the
cell outlet.
•
Semi-micro cell flow rate selected when standard cell is installed
For operation at less than 1.0 mL/min, install a semi-micro flow cell.
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...