Dionex PDA Operator’s Manual
42
Doc. 065378-01 1/12
Factory Default Settings
The Dionex PDA outputs four analog signals at all times, even when the
detector is not interfaced to a computer. Data is presented at the analog
output channels whenever the detector power is on. A chart recorder or
other voltage-measuring device can be used to monitor the channels.
When the Dionex PDA power is first turned on, the analog output settings
are at the factory defaults listed in the table below.
These default settings are also in effect when the power is first turned on:
•
Rise time: 2.0 seconds
•
UV lamp: On
•
Visible lamp: On
Stand-Alone Operation
The table below indicates which Dionex PDA parameter each TTL input
controls during stand-alone operation.
Channel
Single
Wavelength
Bandwidth
Reference
Wavelength
Full-Scale
Output
% Offset
1
210
10
Off
1.0 AU
10%
2
254
10
Off
1.0 AU
10%
3
272
10
Off
1.0 AU
10%
4
520
10
Off
1.0 AU
10%
TTL Input
Dionex PDA Parame-
ter
TTL 1 in
Autozero
TTL 2 in
Reserved
TTL 3 in
Deuterium lamp on/off
TTL 4 in
Tungsten lamp on/off
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...