Dionex PDA Operator’s Manual
16
Doc. 065378-01 1/12
2.3.1
Flow Cells
lists the flow cells available for use with the Dionex PDA.
STANDARD FLOW CELLS
Cell Material
Cell
Volume
Cell Path
Length
Volume of Heat
Exc
Inlet Tubing
Intended Flow
Rate Range
PEEK
(P/N 056346)
13
P
L
10 mm
45
P
L
0.5 to 5.0 mL/min
316 stainless steel
(P/N 056126)
13
P
L
10 mm
20
P
L
0.5 to 5.0 mL/min
SEMI-MICRO FLOW CELLS
Cell Material
Cell
Volume
Cell Path
Length
Volume of Heat
Exc
Inlet Tubing
Intended Flow
Rate Range
PEEK
(P/N 064169)
3.1
P
L
9 mm
5
P
L
Up to 1.0 mL/min
316 stainless steel
(P/N 064168)
3.1
P
L
9 mm
5
P
L
Up to 1.0 mL/min
SEMI-PREPARATIVE FLOW CELL
Cell Material
Cell
Volume
Cell Path
Length
Volume of Heat
Exc
Inlet Tubing
Intended Flow
Rate Range
PEEK
(P/N 064167)
0.7
P
L
0.4 mm
N/A
5.0 to
100.0 mL/min
Table 2-1. Dionex PDA Flow Cells
Do not touch the cell windows. If you touch a window, clean it with
isopropyl alcohol (IPA) and a clean lens tissue.
Strong bases can etch the fused silica windows of the flow cell. If the
mobile phase is a base, make sure the mobile phase concentration
does not exceed 0.1 M. If the concentration of the base is greater than
50 mM, disconnect the separator column and flush the system with
deionized water for 5 minutes at 1.0 mL/min immediately after the
analysis. If strong base remains in the cell for 1 to 2 days, the cell
windows may need to be replaced (see
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...