2 • Description
Doc. 065378-01 1/12
15
Optical System
Component
Function
Tungsten Lamp
The light source for visible and near-infrared wavelengths (380 to
800 nm). The tungsten lamp is focused through an opening in the
internal structure of the deuterium lamp; this allows the lamps to
share the optical axis to the source lens.
Visible Focus Lens
Focuses the visible light from the tungsten lamp to the through-
hole of the deuterium lamp.
Deuterium Lamp
The light source for UV wavelengths (190 to 380 nm).
Source Lens
Receives the light from the tungsten and deuterium lamps and
focuses it so that the beam passes through the flow cell.
Flow Cell
The measurement site for sample absorbance. The cell has a flat
window on each side. Thermo Fisher Scientific offers five cells
for use with the Dionex PDA: standard cells (in both PEEK and
316 stainless steel), semi-micro cells (in both PEEK and 316
stainless steel), and semi-preparative cells (in PEEK only).
Spectrograph Lens
Receives the light from the flow cell and focuses it onto the slit.
Filter Paddle
The motorized filter paddle in the spectrograph has three
positions:
•
Open
= Light passes, unobstructed, along the light path when
the holmium oxide filter is moved out of the light path.
•
Blocked
= No light passes along the light path; used to
measure the dark signal without turning off the lamps.
•
Holmium
= Places the holmium oxide filter in the light path;
used to verify wavelength accuracy.
Slit
The width of the slit (equivalent to 1 nm) optimizes the optical
resolution.
Grating
Diffracts the light beam into its component wavelengths and
directs the light onto the photodiode array.
Photodiode Array
A series of 1024 photosensitive elements. Each element measures
a narrow band of the spectrum. A spectrum is obtained by
measuring the light intensity of each wavelength and reporting the
results over the selected wavelength range.
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...