Dionex PDA Operator’s Manual
112
Doc. 065378-01 1/12
•
The Dionex PDA can be positioned in many locations within a system; the
options include:
Above a Dionex ICS-5000 Thermal Compartment
Below a Thermo Scientific Dionex AS-AP Autosampler
On either side of an integrated system (for example, the Dionex ICS-2100 Ion
Chromatography System)
•
Optical detectors are sensitive to changes in temperature and light. Protect the
Dionex PDA, the columns, and all tubing connections from drafts. Do not
operate the Dionex PDA unless the front cover, lamp cover, and flow cell
cover are in place.
B.2 Unpacking
•
Carefully remove the Dionex PDA from the shipping container.
•
Do not turn on the Dionex PDA unless it is at room temperature (4 to 40 ºC;
40 to 104 ºF). If you move the detector from a cold environment to a warm
environment, wait at least 1 hour for condensation to evaporate before turning
on the power.
•
Unpack all items in the Dionex PDA Ship Kit (P/N 060977) and check them
against the packing list. If there are any discrepancies, notify Thermo Fisher
Scientific immediately.
NOTE Keep the original shipping container and all packing
material. These will be needed if the detector is ever
shipped or is moved to a new location.
Lift the Dionex PDA by the side or bottom only; the front cover of the
detector snaps into place and must not be used for lifting.
Содержание 074114
Страница 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...
Страница 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...
Страница 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...
Страница 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...
Страница 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...
Страница 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...
Страница 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...
Страница 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...