Functional Description
_________________________________________________________________ API Source
__________________ LCQ Deca XP Plus Hardware Manual __________________
1-3
The ESI slide assembly allows the probe to move parallel to the axis of the ion
transfer capillary, or orthogonal to the axis of the ion transfer capillary. A
probe retainer bolt can be used to secure the ESI probe to the slide of the ESI
slide assembly, and prevent movement of the probe orthogonal to the axis of
the ion transfer capillary. Even when the probe retainer bolt is in use for this
purpose, the probe can be moved parallel to the axis of the ion transfer
capillary.
Figure 1-1.
ESI probe assembly, showing AUX gas connected
Flange
Retainer Bolt
Probe Retainer
Bolt
Auxiliary Gas
Line Fitting
(Green)
Probe Slide
Adapter
Probe
Position
Adjustment
Screw
Sheath Gas
Line Fitting
(Blue)
Grounded
Fitting
Holder
ESI
Probe
ESI Needle
High Voltage
Connector
Sample Inlet
(Red)