Safety Precautions
Observe the following safety precautions when you operate or perform
service on the MS detector.
Do Not Perform Any Servicing Other Than That Contained in the
LCQ Deca XP Plus Hardware Manual.
To avoid personal injury or damage to the instrument, do not perform any
servicing other than that contained in the LCQ Deca XP Plus Hardware
Manual or related manuals unless you are qualified to do so.
Shut Down the MS Detector and Disconnect It From Line Power Before
You Service It.
High voltages capable of causing personal injury are used in the instrument.
Some maintenance procedures require that the MS detector be shut down and
disconnected from line power before service is performed. Do not operate the
MS detector with the top or side covers off. Do not remove protective covers
from PCBs.
Respect Heated Zones.
Treat heated zones with respect. The ion transfer capillary and the APCI
vaporizer may be very hot and may cause severe burns if they are touched.
Allow heated components to cool before you service them.
Place the MS Detector in Standby (or Off) Before You Open the
Atmospheric Pressure Ionization (API) Source.
The presence of atmospheric oxygen in the API source when the MS detector
is On could be unsafe. (LCQ Deca XP Plus automatically turns the MS
detector Off when you open the API source, however, it is best to take this
added precaution.)
Make Sure You Have Sufficient Nitrogen For Your API Source.
Before you begin normal operation each day, make sure that you have
sufficient nitrogen for your API source. The presence of atmospheric oxygen
in the API source when the MS detector is On could be unsafe. (LCQ Deca
XP Plus automatically turns the MS detector Off when you run out of
nitrogen, however, it is best to take this added precaution.)
Provide an Adequate Fume Exhaust System and Contain Waste Streams.
It is your responsibility to provide an adequate fume exhaust system. Samples
and solvents that are introduced into the LCQ Deca XP Plus will eventually
be exhausted from the rotary-vane pump. Therefore, the rotary-vane pump
should be connected to a fume exhaust system. Consult local regulations for
the proper method of exhausting the fumes from your system.