Service Maintenance
__________________________________________ Purging the Oil in the Rotary-Vane Pump
__________________ LCQ Deca XP Plus Hardware Manual _________________
5-65
5.11 Purging the Oil in the Rotary-Vane
Pump
You need to purge (decontaminate) the oil in the rotary-vane pump on a daily
basis to remove water and other dissolved chemicals from the pump oil.
Water and other chemicals in the rotary-vane pump can cause corrosion and
decrease the lifetime of the pump. A good time to purge the oil is at the end of
the working week after you flush the API probe, spray shield, and ion transfer
capillary.
To purge the oil in the rotary-vane pump, proceed as follows:
1. Turn off the flow of sample solution from the LC to the MS detector.
2. From the Tune Plus window, choose
Control | Standby
(or click on the
On/Standby button) to put the MS detector in Standby.
3. Withdraw the API flange from the spray shield and place a septum over
the entrance to the ion transfer capillary.
4. Open the gas ballast valve on the rotary-vane pump by turning it to
position
|
. Refer to the manual that came with the pump for the location
of the gas ballast valve.
5. Allow the pump to run for 2 hours with the gas ballast valve open.
6. After 2 hours, close the gas ballast valve by turning it to position O.
7. Top off the oil in the rotary-vane pump reservoir as needed (when the
level of the oil drops below two-thirds of the maximum level visible in the
viewing window).