Страница 1: ......
Страница 2: ...ethods 18 8 Battery 20 9 Readjustments alterations and repairs 21 10 Cleaning and maintenance 21 11 Storage 22 12 Technical checks 22 13 Troubleshooting 24 14 Disposal of the Unit 25 15 Warranty perio...
Страница 3: ...rical conductive medium The irritation can usually be reduced by using an alternate conductive medium like gel or alternate electrode placement 7 Therapeutix TENS PMS must be kept out of the reach of...
Страница 4: ...ription of theDevice 6 1 General Description of theDevice Therapeutix TENS PMS is a portable and DC 3 7V battery powered multifunction device offering both Transcutaneous Electrical Nerve Stimulation...
Страница 5: ...res A large easyto read LCD display Adjustable timer 6 Modes Rechargeable lithium battery 6 4 Mode Description We suggested that you initially experiment using each of the 6 modes The mode that gives...
Страница 6: ...is cycle all the time Modes Parameters Graphical Description Mode 1 1 Timing 10 60 minutes 2 Frequency 68Hz 3 Pulse width 100 S 4 On time 3s 5 Off time 1 56s 6 Duty cycle 4 56s The amplitude for the w...
Страница 7: ...e by 10 minutes After the time runs out the device turns off automatically and it can be restarted if treatment is needed to be continued 9 Sit back and enjoythe deep soothing sensations Notes The The...
Страница 8: ...1 Twin mode or you may place one electrode on the painful site and the other near the site of pain 2 Intensity The intensity can be gradually increased up to the point when it becomes uncomfortable A...
Страница 9: ...in in Neck Figure 3 Mode 1 A for 10 20 minutes and Mode 3 for 10 20 minutes twice or 3 times a day Keep the neck warm and avoid sleeping on a high pillow Pain in shoulders Mode 4 B C for 10 20 minutes...
Страница 10: ...ntended to stimulate healthy muscles in order to improve or facilitate muscle performance It is not intended as therapy for any medical condition Users can choose Mode 1 2 3 or 6 to stimulate the foll...
Страница 11: ...device needscharging Recharge it and then continue to use the device Do not use the device while charging 9 Readjustments alterations and repairs 1 Do not disassemble repair or modify the device with...
Страница 12: ...lowing this Instruction for Use The expired electrodes are to be recycled and do not harm environment Warning The electrodes are intended for single patient useonly 11 Storage Caution Do not store in...
Страница 13: ...een portable and mobile RF communications equipment and the TENS unit Problems Possible causes Try thissolution No One pad feels stronger than the other This is normal Different areas of your bodywill...
Страница 14: ...e expected effect If this equipment is found to cause or respond to interference attempt to correct the problem by one or more of the following measures If feeling too weak or too strong stimulation a...
Страница 15: ...apeutix TENS PMS is used exceeds the applicable RF compliance level above the Therapeutix TENS PMS should be observed to verify normal operation If abnormal performance is observed additional measures...
Страница 16: ...output power of transmitter W Separation distance according to frequency of transmitter m 150 KHz to 80 MHz d 1 167 P 80 MHz to 800 MHz d 1 167 P 800 MHz to 2 5 GHz d 2 333 P 0 01 0 117 0 117 0 233 0...