Cleaning and maintenance
Before cleaning THERA-fit plus always
pull out the net adaptation.
Use only soft damp cloths for cleaning and di-
sinfecting the THERA-fit.
Methylated spirit or conventional methanol-
free, alcohol-based disinfectants can be used
for disinfection. Information on disinfectants
can be found, for instance, in the list of tested
and recognised disinfectants published by the
Robert-Koch-Institut.
The following substances must not
be used:
❑
Trichloroethylene
❑
Acetone
❑
Butanone
❑
Benzol
❑
Methanol
❑
Nitro thinners
❑
1-propanol
❑
2-propanol
Cleaning the THERA-fit
When cleaning and disinfecting
the exerciser, use only soft, damp
(not wet) cloths. Do not spray dis-
infectant!
The exerciser is maintenance-free.
Warranty Battery and fuse
change
Battery change
If the light diodes on the
remote control 8
blink after pressing the ON/OFF button or are
only weakly illuminated you have to change
the battery in the remote control
8
. Pull the
housing on the backside of the remote control
down and take it off. Take out the empty bat-
tery and detach the cable-clip. Set in the new
battery in reversed order.
The battery for the display
2
is on the
lower side of the device in the battery
box. You have to change it as soon as the dis-
play is no more readable.
Open the battery box cap e. g. with a screw
driver. Make sure the right polarity when set-
ting the new battery in.
Remark:
There are two batteries in the bat-
tery box, but there is only one needed for the
function of the training computer. The lower
battery is reserve.
Changing fuse
Changing the fuse should be done by a pro-
fessional dealer.
Warranty
You get a 12 months warranty on THERA-fit
which sets in at the date of delivery, in other
words the moment the bill has been made.
medica Medizintechnik GmbH ensures you to
replace faulty parts of the device within this
time free of charge or to repair the device free
of charge in our factory or in a factory ac-
cepted by medica Medizintechnik GmbH.
There is no warranty on worn material.
Also, we cannot give any warranty on devices
that have been opened, for other parts that
have been added or built in or if repairs have
been made by people not being accepted by
medica Medizintechnik GmbH. There is also
no warranty on any damage caused as a re-
sult of not having observed the regulations
given.