84
4.8.34 MANU:UTILity:GROUNDMODE
Set
Query
Description
Sets or returns the Grounding mode of the current test.
The Ground Mode setting cannot be turned on with the IR and
GB function.
Syntax
Query Syntax
MANU:UTILity:GROUNDMODE {ON|OFF}
MANU:UTILity:GROUNDMODE?
Parameter/ Return
parameter
OFF
Turns ground mode off.
ON
Turns ground mode on.
Example
MANU:UTIL:GROUNDMODE ON
Turns GROUND MODE on.
4.8.35 MANU<x>:EDIT:SHOW
Query
Description
Returns the test parameters of a manual test.
Query Syntax
MANU<x>:EDIT:SHOW?
Parameter
<x>
<NR1> 000~100. Manual test number
Return parameter <string>
Returns a string in the following format:
Test function, test voltage, HI SET value, LO
SET value, Ramp time, test time.
Example
MANU1:EDIT:SHOW ?
>ACW,0.100kV,H=01.00mA,L=00.00mA,R=000.1S,T=001.0S
Returns the test parameters of manual test number 1.
4.9 Sweep Commands
4.9.1 SWEEP:DATA:STATus
Query
Description
Returns the sweep mode, the voltage and current settings
and the number data points that are used in the last sweep.
There can be a maximum of 190 data points, depending on
the testing time.
The data is returned as a string in the following format:
SWEEP MODE,VSET,ISET,Get Data[#data points].
Query Syntax
SWEEP:DATA:STATus?
Return parameter <string> SWEEP MODE, VSET+unit, ISET+units, Get
Data=number of data points
Example
SWEEP:DATA:STATus?
>ACW,V=0.108kV,HI=10.96 mA ,Get Data=011