82
Syntax
Query Syntax
MANU:GB:TTIMe <NR2>
MANU:GB:TTIMe?
Parameter/ Return
parameter
<NR2>
0.5 ~ 999.9 seconds
Example
MANU:GB:TTIM 1
Sets the GB test time to 1 second.
4.8.27 MANU:GB:FREQuency
Set
Query
Description
Sets or returns the GB test frequency in Hz. The test must first
be in GB mode before this command can be used.
Syntax
Query Syntax
MANU:GB:FREQuency {50|60}
MANU:GB:FREQuency?
Parameter/ Return
parameter
50
50 Hz
60
60 Hz
Example
MANU:GB:FREQ 50
Sets the GB test frequency to 50Hz.
4.8.28 MANU:GB:REF
Set
Query
Description
Sets or returns the GB reference value in mΩ. The test must
first be in GB mode before this command can be used.
The GB reference value must be less than the HI SET value.
Syntax
Query Syntax
MANU:GB:REF <NR2>
MANU:GB:REF?
Parameter/ Return
parameter
<NR2>
0.000 ~ 649.9
Example
MANU:GB:REF 100
Sets the GB reference to 100 mΩ.
4.8.29 MANU:GB:ZEROCHECK
Set
Query
Description
Performs the zero check function. The test must first be in
GB mode and in the Ready Status before this command can
be used.
See page 39 for details on the ZERO function.
Syntax
Query Syntax
MANU:GB:ZEROCHECK {ON|OFF}
MANU:GB:ZEROCHECK?
Parameter/ Return
parameter
ON
Zero function is active.
OFF
Zero function is not active.
Example
MANU:GB:ZEROCHECK OFF
Activates the ZERO function.